Determination of strontium 90 in environmental samples by triple quadrupole ICP-MS and its application to Fukushima soil samples

The determination of 90 Sr by ICP-MS is capable of providing a high sample throughput compared with radiometric analytical methods. Although the isobaric interference caused by 90 Zr ions can be reduced by reaction with oxygen, the high background caused by stable 88 Sr has made it difficult to dete...

Full description

Saved in:
Bibliographic Details
Published inJournal of analytical atomic spectrometry Vol. 33; no. 6; pp. 181 - 185
Main Authors Ohno, Takeshi, Hirono, Mutsumi, Kakuta, Shinichiro, Sakata, Shuhei
Format Journal Article
LanguageEnglish
Published London Royal Society of Chemistry 01.01.2018
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The determination of 90 Sr by ICP-MS is capable of providing a high sample throughput compared with radiometric analytical methods. Although the isobaric interference caused by 90 Zr ions can be reduced by reaction with oxygen, the high background caused by stable 88 Sr has made it difficult to determine 90 Sr by conventional ICP-MS. In this study, a rapid and sensitive method was developed for the determination of 90 Sr using triple quadrupole ICP-MS. The MS/MS configuration improves the abundance sensitivity by more than two orders of magnitude (5 × 10 −12 ), and the detection limit for 90 Sr in 1 g environmental samples was 0.6 Bq kg −1 . The applicability of the method to the measurement of 90 Sr in environmental samples was demonstrated by analysis of certified reference materials (soil: IAEA 375; plant: IAEA 330, IAEA 373) contaminated by the Chernobyl accident. The measured 90 Sr radioactivities in the samples were consistent with values determined by radiometric analysis, within analytical error. The measured 90 Sr concentration in IAEA 372 was 610 ± 40 Bq kg −1 (SD, n = 3) (Decay correction was made to December 1, 2016). This method provides a powerful tool for the investigation of radiostrontium contaminations in the Fukushima area and elsewhere. The determination of 90 Sr by triple quadrupole ICP-MS is capable of providing a high sample throughput compared with radiometric analytical methods.
ISSN:0267-9477
1364-5544
DOI:10.1039/c8ja00017d