Low Limit for Effective Signal Strength in the Stratus OCT in Imperative Low Signal Strength Cases
To determine the lowest limit of signal strength that is still effective for accurate analysis of optic coherence tomography (OCT) values, we investigated the reproducibility of OCT scans by signal strength (SS). A total of 668 subjects were scanned for measurements of retinal nerve fiber layer (RNF...
Saved in:
Published in | Korean journal of ophthalmology Vol. 26; no. 3; pp. 182 - 188 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
Korea (South)
The Korean Ophthalmological Society
01.06.2012
대한안과학회 |
Subjects | |
Online Access | Get full text |
ISSN | 1011-8942 2092-9382 2092-9382 |
DOI | 10.3341/kjo.2012.26.3.182 |
Cover
Summary: | To determine the lowest limit of signal strength that is still effective for accurate analysis of optic coherence tomography (OCT) values, we investigated the reproducibility of OCT scans by signal strength (SS).
A total of 668 subjects were scanned for measurements of retinal nerve fiber layer (RNFL) thickness using the Stratus OCT twice on the same day. The variability of overall RNFL thickness parameters obtained at different SS was analyzed and compared by repeated-measures of ANOVA and Spearman's correlation coefficient. Values of the intraclass correlation coefficient (ICC) and variability (standard deviation) of RNFL thickness were obtained. The false positive ratio was analyzed.
When SS was 3, the variability of RNFL thickness was significantly different (low ICC, high variability) in comparison to when SS was 4 or greater. Significant negative correlations were observed between variability in RNFL thickness and signal strength. The difference of variability of average RNFL thickness between SS 4 (4.94 µm) and SS 6 (4.41 µm) was 0.53 µm.
Clinically, the difference of variability of average RNFL thickness between SS 4 and SS 6 was quite small. High SS is important, however, when signal strength is low due to uncorrectable factors in patients in need of OCT for glaucoma and retinal disease. Our results suggest that SS 4 is the lowest acceptable limit of signal strength for obtaining reproducible scanning images. |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 G704-002170.2012.26.3.012 |
ISSN: | 1011-8942 2092-9382 2092-9382 |
DOI: | 10.3341/kjo.2012.26.3.182 |