MR mammographic localization. Work in progress

The authors present phantom test results for a stereotaxic device that may permit simple, rapid, and accurate needle biopsy and localization of breast lesions detected at contrast material-enhanced magnetic resonance (MR) mammography. The mechanical accuracy of this prototype MR breast localizer is...

Full description

Saved in:
Bibliographic Details
Published inRadiology Vol. 189; no. 3; p. 915
Main Authors Hussman, K, Renslo, R, Phillips, J J, Fischer, H J, Khalkhali, I, Braslau, D L, Sinow, R M
Format Journal Article
LanguageEnglish
Published United States 01.12.1993
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:The authors present phantom test results for a stereotaxic device that may permit simple, rapid, and accurate needle biopsy and localization of breast lesions detected at contrast material-enhanced magnetic resonance (MR) mammography. The mechanical accuracy of this prototype MR breast localizer is approximately plus or minus 3.5 mm at 5 cm.