A new permeance meter based on both lumped elements/transmission line theories

We have developed a new permeance meter in the 1 MHz-1.5 GHz range using either a microstrip pickup coil or a shielded loop pickup coil. Both pickup coils and the driving plates are free from LC resonance due to impedance matching. In the MHz range we must consider signal-to-noise ratio of the picku...

Full description

Saved in:
Bibliographic Details
Published inIEEE transactions on magnetics Vol. 32; no. 5; pp. 4941 - 4943
Main Authors Yamaguchi, M., Yabukami, S., Arai, K.I.
Format Journal Article
LanguageEnglish
Published IEEE 01.09.1996
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:We have developed a new permeance meter in the 1 MHz-1.5 GHz range using either a microstrip pickup coil or a shielded loop pickup coil. Both pickup coils and the driving plates are free from LC resonance due to impedance matching. In the MHz range we must consider signal-to-noise ratio of the pickup coil. In the GHz range we must eliminate the voltage induced by the electric field, and an electromagnetic higher order mode in the driving plates.
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ISSN:0018-9464
1941-0069
DOI:10.1109/20.539295