A new permeance meter based on both lumped elements/transmission line theories
We have developed a new permeance meter in the 1 MHz-1.5 GHz range using either a microstrip pickup coil or a shielded loop pickup coil. Both pickup coils and the driving plates are free from LC resonance due to impedance matching. In the MHz range we must consider signal-to-noise ratio of the picku...
Saved in:
Published in | IEEE transactions on magnetics Vol. 32; no. 5; pp. 4941 - 4943 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
IEEE
01.09.1996
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | We have developed a new permeance meter in the 1 MHz-1.5 GHz range using either a microstrip pickup coil or a shielded loop pickup coil. Both pickup coils and the driving plates are free from LC resonance due to impedance matching. In the MHz range we must consider signal-to-noise ratio of the pickup coil. In the GHz range we must eliminate the voltage induced by the electric field, and an electromagnetic higher order mode in the driving plates. |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Feature-1 content type line 23 |
ISSN: | 0018-9464 1941-0069 |
DOI: | 10.1109/20.539295 |