Metronome-guided cochlear implantation for slower and smoother insertions of lateral wall electrodes
Introduction Achieving a slow and smooth electrode array insertion is paramount for preserving structural and functional integrity during cochlear implantation. This controlled study evaluates the efficacy of a metronome-guided insertion technique in enhancing the smoothness and speed of electrode a...
Saved in:
Published in | European archives of oto-rhino-laryngology Vol. 281; no. 9; pp. 4603 - 4609 |
---|---|
Main Authors | , , , , , , , |
Format | Journal Article |
Language | English |
Published |
Berlin/Heidelberg
Springer Berlin Heidelberg
01.09.2024
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Introduction
Achieving a slow and smooth electrode array insertion is paramount for preserving structural and functional integrity during cochlear implantation. This controlled study evaluates the efficacy of a metronome-guided insertion technique in enhancing the smoothness and speed of electrode array insertions.
Methods
In a prospective cohort study, patients undergoing cochlear implant surgery between 2022 and 2023 with lateral wall electrode arrays were included. Metronome guidance was delivered through an acoustic signal via headphones during electrode array insertion in cochlear implantation and compared to a control group without metronome-guidance.
Results
In total, 37 cases were evaluated, including 25 conventional insertions and 12 metronome-guided insertions. The results indicate that metronome-guided insertions were significantly slower (− 0.46 mm/s;
p
< 0.001) without extending the overall procedure time. This can be attributed to fewer paused sections observed in the metronome-guided technique. Moreover, metronome-guided insertions exhibited superior performance in terms of insertion smoothness and a reduced number of re-gripping events.
Conclusions
The findings support the recommendation for the systematic application of metronome guidance in the manual insertion of cochlear implant electrode arrays, emphasizing its potential to optimize surgical outcomes. |
---|---|