Real-world insight into public access defibrillator use over five years

Public access defibrillators (PADs) represent unique life-saving medical devices as they may be used by untrained lay rescuers. Collecting representative clinical data on these devices can be challenging. Here, we present results from a retrospective observational cohort study, describing real-world...

Full description

Saved in:
Bibliographic Details
Published inOpen heart Vol. 7; no. 1; p. e001251
Main Authors Torney, Hannah, McAlister, Olibhéar, Harvey, Adam, Kernaghan, Amy, Funston, Rebecca, McCartney, Ben, Davis, Laura, Bond, Raymond, McEneaney, David, Adgey, Jennifer
Format Journal Article
LanguageEnglish
Published England BMJ Publishing Group 01.06.2020
SeriesOriginal research
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Public access defibrillators (PADs) represent unique life-saving medical devices as they may be used by untrained lay rescuers. Collecting representative clinical data on these devices can be challenging. Here, we present results from a retrospective observational cohort study, describing real-world PAD utilisation over a 5-year period. Data were collected between October 2012 and October 2017. Responders voluntarily submitted electronic data downloaded from HeartSine PADs, and patient demographics and other details using a case report form in exchange for a replacement battery and electrode pack. Data were collected for 977 patients (692 males, 70.8%; 255 females, 26.1%; 30 unknown, 3.1%). The mean age (SD) was 59 (18) years (range <1 year to 101 years). PAD usage occurred most commonly in homes (n=328, 33.6%), followed by public places (n=307, 31.4%) and medical facilities (n=128, 13.1%). Location was unknown in 40 (4.09%) events. Shocks were delivered to 354 patients. First shock success was 312 of 350 patients where it could be determined (89.1%, 95% CI 85.4% to 92.2%). Patients with reported response times ≤5 min were more likely to survive to hospital admission (89/296 (30.1%) vs 40/250 (16.0%), p<0.001). Response time was unknown for 431 events. This is the first study to report global PAD usage in voluntarily submitted, unselected real-world cases and demonstrates the real-world effectiveness of PADs, as confirmed by first shock success.
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Undefined-1
ObjectType-Feature-3
content type line 23
ISSN:2053-3624
2053-3624
DOI:10.1136/openhrt-2020-001251