Reliability and quality of water isotope data collected with a low-budget rain collector
RATIONALE Low‐budget rain collectors for water isotope analysis, such as the 'ball‐in‐funnel type collector' (BiFC), are widely used in studies on stable water isotopes of rain. To date, however, an experimental quality assessment of such devices in relation to climatic factors does not ex...
Saved in:
Published in | Rapid communications in mass spectrometry Vol. 28; no. 8; pp. 879 - 885 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
England
Blackwell Publishing Ltd
30.04.2014
Wiley Subscription Services, Inc |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | RATIONALE
Low‐budget rain collectors for water isotope analysis, such as the 'ball‐in‐funnel type collector' (BiFC), are widely used in studies on stable water isotopes of rain. To date, however, an experimental quality assessment of such devices in relation to climatic factors does not exist.
METHODS
We used Cavity Ring‐Down Spectrometry (CRDS) to quantify the effects of evaporation on the δ18O values of reference water under controlled conditions as a function of the elapsed time between rainfall and collection for isotope analysis, the sample volume and the relative humidity (RH: 31% and 67%; 25°C). The climate chamber conditions were chosen to reflect the warm and dry end of field conditions that favor evaporative enrichment (EE). We also tested the performance of the BiFC in the field, and compared our δ2H/δ18O data obtained by isotope ratio mass spectrometry (IRMS) with those from the Swiss National Network for the Observation of Isotopes in the Water Cycle (ISOT).
RESULTS
The EE increased with time, with a 1‰ increase in the δ18O values after 10 days (RH: 25%; 25°C; 35 mL (corresponding to a 5 mm rain event); p <0.001). The sample volume strongly affected the EE (max. value +1.5‰ for 7 mL samples (i.e., 1 mm rain events) after 72 h at 31% and 67% RH; p <0.001), whereas the relative humidity had no significant effect. Using the BiFC in the field, we obtained very tight relationships of the δ2H/δ18O values (r2 ≥0.95) for three sites along an elevational gradient, not significantly different from that of the next ISOT station.
CONCLUSIONS
Since the chosen experimental conditions were extreme compared with the field conditions, it was concluded that the BiFC is a highly reliable and inexpensive collector of rainwater for isotope analysis. Copyright © 2014 John Wiley & Sons, Ltd. |
---|---|
Bibliography: | ark:/67375/WNG-CZMZ906G-F ArticleID:RCM6852 istex:A74B72392F92BD03947835DC97FEE5ADB01D351D ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 ObjectType-Article-2 ObjectType-Feature-1 |
ISSN: | 0951-4198 1097-0231 |
DOI: | 10.1002/rcm.6852 |