Fatigue properties of the hoffmann-vidal external fixation apparatus

Fatigue tests of the standard HoffmannVidal (H-V) external fixator were performed. After 2,500,000 cycles, fractures occurred in the post attached to the pin bracket. The screws on the pin fixation bracket tend to become loose after cyclic loading test, and periodic tightening of all the fixation sc...

Full description

Saved in:
Bibliographic Details
Published inOrthopedics (Thorofare, N.J.) Vol. 7; no. 3; pp. 443 - 456
Main Authors Shiba, R, Chao, E Y, Kasman, R
Format Journal Article
LanguageEnglish
Published United States SLACK INCORPORATED 01.03.1984
Online AccessGet full text

Cover

Loading…
More Information
Summary:Fatigue tests of the standard HoffmannVidal (H-V) external fixator were performed. After 2,500,000 cycles, fractures occurred in the post attached to the pin bracket. The screws on the pin fixation bracket tend to become loose after cyclic loading test, and periodic tightening of all the fixation screws on a H-V apparatus is recommended under in vivo application. In repeated tightening of individual joints, screws on the pin brackets and adjustable connecting rods fractured. Deformation of the retaining set screws in the universal ball joints caused joint stiffness reduction. Severe abrasive wear of the joint surfaces was observed after repeated tightening. It is concluded that this apparatus may be safely used for four to five six-month clinical applications, provided that the critical parts are replaced between sequential applications. Suggestions for clinical management and recommendations for component modifications are made.
Bibliography:SourceType-Other Sources-1
content type line 63
ObjectType-Editorial-2
ObjectType-Commentary-1
ISSN:0147-7447
1938-2367
DOI:10.3928/0147-7447-19840301-11