Improvement of Analysis Precision of Laser Microprobe Analyser by Correcting Self-Absorption

A calculation procedure for correcting self-absorption of the spectral line by using a simple source model is proposed to improve the sensitivity of laser microprobe analysis. A linear calibration curve with a large slope is obtained by this method. This method is found useful for improving the anal...

Full description

Saved in:
Bibliographic Details
Published inJournal of the Spectroscopical Society of Japan Vol. 36; no. 2; pp. 123 - 125
Main Authors KUZUYA, Mikio, MIKAMI, Takayoshi, MIKAMI, Osamu
Format Journal Article
LanguageEnglish
Published The Spectroscopical Society of Japan 1987
Online AccessGet full text

Cover

Loading…
More Information
Summary:A calculation procedure for correcting self-absorption of the spectral line by using a simple source model is proposed to improve the sensitivity of laser microprobe analysis. A linear calibration curve with a large slope is obtained by this method. This method is found useful for improving the analysis precision of the laser microprobe analyser.
ISSN:0038-7002
1884-6785
DOI:10.5111/bunkou.36.123