Dry heat loss in incubator: comparison of two premature newborn sized manikins
Keeping premature newborns warm is crucial for their survival. Their ability to prevent excessive heat loss to the environment and to control their body temperature is limited. The risk of hypothermia is particularly important for low-birth-weight newborns with a large body surface area in relation...
Saved in:
Published in | European journal of applied physiology Vol. 92; no. 6; pp. 679 - 682 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Germany
Springer Nature B.V
01.09.2004
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Keeping premature newborns warm is crucial for their survival. Their ability to prevent excessive heat loss to the environment and to control their body temperature is limited. The risk of hypothermia is particularly important for low-birth-weight newborns with a large body surface area in relation to their mass of heat-producing tissues. The present study was performed to assess the body heat loss difference between small and large body-size premature newborns using two anthropomorphic thermal manikins of premature newborns of 900 g and 1,800 g (respective body surface areas of 0.086 and 0.150 m2). The dry heat loss from the six body segments of the small manikin (S) was measured and compared with that of the large manikin (L). The two manikins were exposed to five different environmental temperatures ranging between 29 and 35 degrees C in a single-walled, air-heated closed incubator. The magnitudes of heat loss decreased significantly by 20.4% between the two manikins [small manikin 110.1 (44.3) W/m2 vs large manikin 87.6 (25.8) W/m2, mean values with one standard deviation]. The results obtained from the comparison of the heat loss measures from the two manikins confirm the fact that the heat loss increases with an increase in the ratio of the body surface area to body mass. The thermal manikin appears to provide an accurate method for the assessment of thermal conditions in neonatal care. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1439-6319 1439-6327 |
DOI: | 10.1007/s00421-004-1130-5 |