Effects of dextran on five biuret-based procedures for total protein in serum

We evaluated the effect of dextran on values for total protein in serum as measured by the biuret method with five widely used automated instruments: the American Monitor Parallel; the Du Pont aca II; the Roche Cobas-Bio; the Kodak Ektachem 400; and the Beckman Astra 8. Dextran concentrations as gre...

Full description

Saved in:
Bibliographic Details
Published inClinical chemistry (Baltimore, Md.) Vol. 31; no. 12; pp. 2018 - 2019
Main Authors Barnes, DB, Pierce, GF, Lichti, D, Landt, M, Koenig, J, Chan, KM
Format Journal Article
LanguageEnglish
Published Washington, DC Am Assoc Clin Chem 01.12.1985
American Association for Clinical Chemistry
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:We evaluated the effect of dextran on values for total protein in serum as measured by the biuret method with five widely used automated instruments: the American Monitor Parallel; the Du Pont aca II; the Roche Cobas-Bio; the Kodak Ektachem 400; and the Beckman Astra 8. Dextran concentrations as great as 25 or 30 g/L had relatively little or no influence on total protein measurements by the latter three instruments. Dextran concentrations exceeding 6 g/L caused falsely low results with the aca, whereas the Parallel gave falsely high results when the dextran concentration exceeded 2 g/L. The aca total protein procedure could be protected from the interference by dextran concentrations up to 30 g/L by injecting 0.4-0.8 mL of ethylene glycol directly into the reagent pack before sampling. However, we could not eliminate the interference with the Parallel procedure by any simple means; we thus recommend that it not be used for measuring total protein in serum samples from patients who are being treated with dextran.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0009-9147
1530-8561
DOI:10.1093/clinchem/31.12.2018