Cross-Seal IDE Trial: Prospective, Multicenter, Single-Arm Study of the Cross-Seal Suture-Mediated Vascular Closure Device System
BACKGROUND: An increasing number of interventional procedures require large-sheath technology (>12F) with a favorable outcome with endovascular rather than open surgical access. However, vascular complications are a limitation for the management of these patients. This trial aimed to determine th...
Saved in:
Published in | Circulation. Cardiovascular interventions Vol. 17; no. 6; p. e013842 |
---|---|
Main Authors | , , , , , , , , , , , , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
Hagerstown, MD
Lippincott Williams & Wilkins
01.06.2024
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | BACKGROUND:
An increasing number of interventional procedures require large-sheath technology (>12F) with a favorable outcome with endovascular rather than open surgical access. However, vascular complications are a limitation for the management of these patients. This trial aimed to determine the effectiveness and safety of the Cross-Seal suture-mediated vascular closure device in obtaining hemostasis at the target limb access site following interventional procedures using 8F to 18F procedural sheaths.
METHODS:
The Cross-Seal IDE trial (Investigational Device Exemption) was a prospective, single-arm, multicenter study in subjects undergoing percutaneous endovascular procedures utilizing 8F to 18F ID procedural sheaths. The primary efficacy end point was time to hemostasis at the target limb access site. The primary safety end point was freedom from major complications of the target limb access site within 30 days post procedure.
RESULTS:
A total of 147 subjects were enrolled between August 9, 2019, and March 12, 2020. Transcatheter aortic valve replacement was performed in 53.7% (79/147) and percutaneous endovascular abdominal/thoracic aortic aneurysm repair in 46.3% (68/147) of subjects. The mean sheath ID was 15.5±1.8 mm. The primary effectiveness end point of time to hemostasis was 0.4±1.4 minutes. An adjunctive intervention was required in 9.2% (13/142) of subjects, of which 2.1% (3/142) were surgical and 5.6% (8/142) endovascular. Technical success was achieved in 92.3% (131/142) of subjects. Freedom from major complications of the target limb access site was 94.3% (83/88).
CONCLUSIONS:
In selected patients undergoing percutaneous endovascular procedures utilizing 8F to 18F ID procedural sheath, Cross-Seal suture-mediated vascular closure device achieved favorable effectiveness and safety in the closure of the large-bore arteriotomy.
REGISTRATION:
URL: https://www.clinicaltrials.gov; Unique identifier: NCT03756558. |
---|---|
Bibliography: | Supplemental Material is available at https://www.ahajournals.org/doi/suppl/10.1161/CIRCINTERVENTIONS.123.013842. For Sources of Funding and Disclosures, see page 530. Correspondence to: Prakash Krishnan, MD, Zena and Michael A. Wiener Cardiovascular Institute, Icahn School of Medicine at Mount Sinai, One Gustave L. Levy Place, P.B. 1030, New York, NY 10029. Email prakash.krishnan@mountsinai.org ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1941-7640 1941-7632 1941-7632 |