Handcrafted cuff manometers do not accurately measure endotracheal tube cuff pressure

To test the agreement between two handcrafted devices and a cuff-specific manometer. The agreement between two handcrafted devices adapted to measure tracheal tube cuff pressure and a cuff-specific manometer was tested on 79 subjects. The cuff pressure was measured with a commercial manometer and wi...

Full description

Saved in:
Bibliographic Details
Published inRevista Brasileira de terapia intensiva Vol. 27; no. 3; pp. 228 - 234
Main Authors Annoni, Raquel, Almeida Junior, Antonio Evanir de
Format Journal Article
LanguageEnglish
Published Brazil Associação Brasileira de Medicina intensiva 01.07.2015
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:To test the agreement between two handcrafted devices and a cuff-specific manometer. The agreement between two handcrafted devices adapted to measure tracheal tube cuff pressure and a cuff-specific manometer was tested on 79 subjects. The cuff pressure was measured with a commercial manometer and with two handcrafted devices (HD) assembled with aneroid sphygmomanometers (HD1 and HD2). The data were compared using Wilcoxon and Spearman tests, the intraclass correlation coefficient (ICC) and limit-of-agreement analysis. Cuff pressures assessed with handcrafted devices were significantly different from commercial device measurements (pressures were higher when measured with HD1 and lower with HD2). The ICCs between the commercial device and HD1 and HD2 were excellent (ICC = 0.8 p < 0.001) and good (ICC = 0.66, p < 0.001), respectively. However, the Bland- Altman plots showed wide limits of agreement between HD1 and HD2 and the commercial device. The handcrafted manometers do not provide accurate cuff pressure measurements when compared to a cuff-specific device and should not be used to replace the commercial cuff manometers in mechanically ventilated patients.
ISSN:0103-507X
1982-4335
DOI:10.5935/0103-507X.20150037