Provoked and Spontaneous High-Frequency, Low-Amplitude, Respirophasic Noise Transients in Patients with Implantable Cardioverter Defibrillators
Respirophasic Noise Transients in ICD Patients. Introduction: Ventricular oversensing (OS) of respirophasic noise transients may cause spurious detections and therapies and pacing inhibition among patients with implantable cardioverter defibrillators (ICDs). The incidence of OS and its relationship...
Saved in:
Published in | Journal of cardiovascular electrophysiology Vol. 12; no. 4; pp. 402 - 410 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Oxford, UK
Blackwell Science Inc
01.04.2001
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Respirophasic Noise Transients in ICD Patients. Introduction: Ventricular oversensing (OS) of respirophasic noise transients may cause spurious detections and therapies and pacing inhibition among patients with implantable cardioverter defibrillators (ICDs). The incidence of OS and its relationship to clinical variables and ICD system design are unknown.
Methods and Results: Three hundred twenty‐nine patients performed provocative respiratory maneuvers at rest during intrinsic rhythm and continuous ventricular pacing. OS resulting in spurious ventricular detections was provoked in 3 (0.9%) of 329 patients during intrinsic rhythm and 34 (10.3%) of 329 during pacing. Noise transients not recognized and marked as sensed events, but visually evident on the local endocardial ventricular electrogram, were provoked in an additional 23 (7.0%) of 329 patients. Multivariate logistic regression identified history of spontaneous OS (P < 0.0005, odds ratio 9.7, 95% confidence interval [CI] 1.9 to 50.0), automatic gain control device (P < 0.0005, odds ratio 5.3, 95% CI 2.6 to 10.8) or integrated bipolar lead (P = 0.05, odds ratio 2.6, 95% CI 1.0 to 7.25), and male gender (P = 0.008, odds ratio 3.7, 95% CI 1.2 to 11.1) as predictive of provocable OS. Spontaneous OS resulting in spurious ventricular detections and therapies occurred in 12 (3.6%) patients during follow‐up. Eleven of 12 spontaneous episodes occurred in male patients during ventricular pacing; 11 of 12 patients had automatic gain control devices and integrated bipolar leads.
Conclusion: OS is commonly provoked in ICD patients during ventricular pacing and may occur spontaneously, causing spurious tachyarrhythmia therapies and pacing inhibition. Differences in the incidence of spontaneous and provoked OS between ICD systems can be explained on the basis of unique features of automatic sensing systems and sensing lead design. |
---|---|
Bibliography: | ark:/67375/WNG-4G077738-8 ArticleID:JCE402 istex:58111FFDD55798F9976D34D14C5F772DBF30DE32 Presented at the 47th Annual Scientific Sessions of the American College of Cardiology, March 1999. ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |