A large area plastic scintillation detector with 4-corner-readout
A 760 mm × 760 mm× 30 mm plastic scintillation detector viewed by photomultiplier tubes (PMTs) from four corners has been developed, and the detector has been tested with cosmic rays and γrays. A position- independent effective time Teff has been found, indicating this detector can be used as a TOF...
Saved in:
Published in | Chinese physics C Vol. 40; no. 5; pp. 41 - 45 |
---|---|
Main Author | |
Format | Journal Article |
Language | English |
Published |
01.05.2016
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A 760 mm × 760 mm× 30 mm plastic scintillation detector viewed by photomultiplier tubes (PMTs) from four corners has been developed, and the detector has been tested with cosmic rays and γrays. A position- independent effective time Teff has been found, indicating this detector can be used as a TOF detector. The hit position can also be reconstructed by the time from the four corners.A. TOF resolution of 236 ps and a position resolution of 48 mm have been achieved, and the detection efficiency has also been investigated. |
---|---|
Bibliography: | A 760 mm × 760 mm× 30 mm plastic scintillation detector viewed by photomultiplier tubes (PMTs) from four corners has been developed, and the detector has been tested with cosmic rays and γrays. A position- independent effective time Teff has been found, indicating this detector can be used as a TOF detector. The hit position can also be reconstructed by the time from the four corners.A. TOF resolution of 236 ps and a position resolution of 48 mm have been achieved, and the detection efficiency has also been investigated. 11-5641/O4 scintillation detector, 4-corner-readout, time resolution, position resolution ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 1674-1137 0254-3052 |
DOI: | 10.1088/1674-1137/40/5/056001 |