Diode laser spectrometer for high-precision measurements
The issues of the development, calibration and putting into operation of a diode laser spectrometer for high-precision measurements are considered. The tests rely on the use of an isolated P20 line of the 00031 - 00001 band of the 12C16O2 molecule. The spectrometer error budget is analysed. A softwa...
Saved in:
Published in | Quantum electronics (Woodbury, N.Y.) Vol. 49; no. 7; pp. 613 - 622 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
Bristol
Kvantovaya Elektronika, Turpion Ltd and IOP Publishing
01.07.2019
IOP Publishing |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The issues of the development, calibration and putting into operation of a diode laser spectrometer for high-precision measurements are considered. The tests rely on the use of an isolated P20 line of the 00031 - 00001 band of the 12C16O2 molecule. The spectrometer error budget is analysed. A software programme for the approximation of a spectral line contour is designed and the procedure is analysed. It is shown that, for a baseline with a relative level of 2.3 × 10−5, it is optimal to use no more than six fitting parameters in the line contour approximation. The measured integral intensity of the analytical line constitutes 5.3865(85) × 10−23 cm per molecule. |
---|---|
ISSN: | 1063-7818 1468-4799 |
DOI: | 10.1070/QEL16776 |