Diode laser spectrometer for high-precision measurements

The issues of the development, calibration and putting into operation of a diode laser spectrometer for high-precision measurements are considered. The tests rely on the use of an isolated P20 line of the 00031 - 00001 band of the 12C16O2 molecule. The spectrometer error budget is analysed. A softwa...

Full description

Saved in:
Bibliographic Details
Published inQuantum electronics (Woodbury, N.Y.) Vol. 49; no. 7; pp. 613 - 622
Main Authors Nadezhdinskii, A.I., Ponurovskii, Ya.Ya
Format Journal Article
LanguageEnglish
Published Bristol Kvantovaya Elektronika, Turpion Ltd and IOP Publishing 01.07.2019
IOP Publishing
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The issues of the development, calibration and putting into operation of a diode laser spectrometer for high-precision measurements are considered. The tests rely on the use of an isolated P20 line of the 00031 - 00001 band of the 12C16O2 molecule. The spectrometer error budget is analysed. A software programme for the approximation of a spectral line contour is designed and the procedure is analysed. It is shown that, for a baseline with a relative level of 2.3 × 10−5, it is optimal to use no more than six fitting parameters in the line contour approximation. The measured integral intensity of the analytical line constitutes 5.3865(85) × 10−23 cm per molecule.
ISSN:1063-7818
1468-4799