Drug-Coated Balloons for Complex Femoropopliteal Lesions

Abstract Objectives The authors sought to investigate the efficacy of a drug-coated balloon (DCB) for treatment of complex femoropopliteal lesions. Background Superiority of DCBs compared with uncoated balloon angioplasty for femoropopliteal interventions has been demonstrated in randomized trials f...

Full description

Saved in:
Bibliographic Details
Published inJACC. Cardiovascular interventions Vol. 9; no. 7; pp. 715 - 724
Main Authors Schmidt, Andrej, MD, Piorkowski, Michael, MD, Görner, Henrik, MD, Steiner, Sabine, MD, MSc, Bausback, Yvonne, MD, Scheinert, Susanne, MD, Banning-Eichenseer, Ursula, PhD, Staab, Holger, MD, Branzan, Daniela, MD, Varcoe, Ramon L., MD, Scheinert, Dierk, MD
Format Journal Article
LanguageEnglish
Published Elsevier Inc 01.04.2016
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Abstract Objectives The authors sought to investigate the efficacy of a drug-coated balloon (DCB) for treatment of complex femoropopliteal lesions. Background Superiority of DCBs compared with uncoated balloon angioplasty for femoropopliteal interventions has been demonstrated in randomized trials for short lesions. Their performance in complex lesions with higher restenosis rates is unclear. Methods Patency, target lesion revascularization (TLR) rate, clinical improvement, and safety endpoints of femoropopliteal lesions in 288 limbs (n = 260) treated with the In.Pact Pacific or Admiral DCB (Medtronic, Minneapolis, Minnesota) were retrospectively analyzed for up to 2 years of follow-up. Predictors of restenosis were identified by logistic regression. Results Lesions were de novo in 51.7%, restenosis in 11.1%, and in-stent restenosis in 37.2%. Mean lesion length was 24.0 ± 10.2 cm, and 65.3% were occluded. Stent implantation was performed in 23.3%. Kaplan Meier estimates of primary patency were 79.2% and 53.7% for all lesions at 1 and 2 years, respectively, whereas freedom from TLR was 85.4% and 68.6%. Primary patency for in-stent restenosis treatment was 76.6% and 48.6%, and freedom from TLR was 83.0% and 58.7% at 1 and 2 years, respectively. Rutherford category improved from a median 3.3 to 1.2 at 1 year, and to 1.1 at 2 years. Major amputation rate was 2.1% at 2 years. No adverse events were thought to be attributable to the coating of the balloon. Conclusions These results suggest that DCB are safe and effective in delaying rather than preventing restenosis in long, complex lesions and restenosis of the femoropopliteal tract. Further studies are recommended to confirm these results.
ISSN:1936-8798
1876-7605
DOI:10.1016/j.jcin.2015.12.267