Efficiency calibration of low-energy electron detectors by means of β-ray emitters
The relative efficiency of a channel electron multiplier for 2–9 keV electrons was determined using the β-spectra of vacuum evaporated 241 Pu and 63 Ni sources. The results agreed within ±0.8% (1 σ). Conversion electron measurements of 57 Co and 169 Yb supported the result. The electron scattering a...
Saved in:
Published in | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Vol. 438; no. 2; pp. 433 - 438 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Elsevier B.V
11.12.1999
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The relative efficiency of a channel electron multiplier for 2–9
keV electrons was determined using the β-spectra of vacuum evaporated
241
Pu
and
63
Ni
sources. The results agreed within ±0.8% (1
σ). Conversion electron measurements of
57
Co
and
169
Yb
supported the result. The electron scattering and energy losses within the sources were taken into account by means of Monte Carlo simulations of individual elastic and inelastic collisions. The detector was calibrated in the same geometry as it works inside the cylindrical mirror analyzer. |
---|---|
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/S0168-9002(99)00875-X |