Efficiency calibration of low-energy electron detectors by means of β-ray emitters

The relative efficiency of a channel electron multiplier for 2–9 keV electrons was determined using the β-spectra of vacuum evaporated 241 Pu and 63 Ni sources. The results agreed within ±0.8% (1 σ). Conversion electron measurements of 57 Co and 169 Yb supported the result. The electron scattering a...

Full description

Saved in:
Bibliographic Details
Published inNuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Vol. 438; no. 2; pp. 433 - 438
Main Authors Špalek, A., Dragoun, O., Ryšavý, M., Dragounová, N.
Format Journal Article
LanguageEnglish
Published Elsevier B.V 11.12.1999
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The relative efficiency of a channel electron multiplier for 2–9 keV electrons was determined using the β-spectra of vacuum evaporated 241 Pu and 63 Ni sources. The results agreed within ±0.8% (1 σ). Conversion electron measurements of 57 Co and 169 Yb supported the result. The electron scattering and energy losses within the sources were taken into account by means of Monte Carlo simulations of individual elastic and inelastic collisions. The detector was calibrated in the same geometry as it works inside the cylindrical mirror analyzer.
ISSN:0168-9002
1872-9576
DOI:10.1016/S0168-9002(99)00875-X