Endocardial electrogram amplitudes and stimulation thresholds in monopolar electrodes: Possible correlations

The possible correlations between unipolar endocardial electrogram amplitudes and voltage threshold stimulation were tested with electrodes having metallic tips of two different areas. The threshold measurements were always studied with the same pulse characteristics. The data were analyzed through...

Full description

Saved in:
Bibliographic Details
Published inJournal of electrocardiology Vol. 10; no. 4; pp. 359 - 366
Main Authors Antonioli, Enrico G., Bagni, Bruno, Baggioni, Franco G., Grassi, Gino
Format Journal Article
LanguageEnglish
Published United States Elsevier Inc 1977
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The possible correlations between unipolar endocardial electrogram amplitudes and voltage threshold stimulation were tested with electrodes having metallic tips of two different areas. The threshold measurements were always studied with the same pulse characteristics. The data were analyzed through five mathematical functions (linear, exponential, potential, hyperbola and translate hyperbola). The results show a significant correlation, mainly by hyperbolic and exponential functions. Consequently, we could deduce a parameter value since we already knew the other one. This was possible not only during the first implantation, but also later on.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0022-0736
1532-8430
DOI:10.1016/S0022-0736(77)80009-5