Method development for the determination of alkali metals in samples from pyrochemical reprocessing using ICP-OES and comparison with sector field ICP-MS

An analytical procedure for the reliable determination of the alkaline elements Na, K, and Li in radioactive samples was developed and validated. To this end, a high efficiency, desolvating nebulizer (Apex E) was coupled to a commercial high resolution ICP-OES instrument operated at 1000W. Because o...

Full description

Saved in:
Bibliographic Details
Published inMicrochemical journal Vol. 105; pp. 9 - 14
Main Authors Krachler, Michael, Van Winckel, Stefaan, Cardinale, Mariangela, Lynch, Brian, Murakami, Tsuyoshi
Format Journal Article
LanguageEnglish
Published Elsevier B.V 01.11.2012
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:An analytical procedure for the reliable determination of the alkaline elements Na, K, and Li in radioactive samples was developed and validated. To this end, a high efficiency, desolvating nebulizer (Apex E) was coupled to a commercial high resolution ICP-OES instrument operated at 1000W. Because of their radioactive nature, all investigated samples were handled in appropriate glove boxes operated in under-pressure, following strict safety and security procedures. After optimization of the ICP-OES operational parameters, detection limits (LODs) of 0.15μg/L, 0.8μg/L, and 1.3μg/L were obtained for Li (λ=670.784nm), Na (λ=589.592nm), and K (λ=766.490nm). Compared to conventional sample uptake rates of 1–2mL/min, the 350μL/min used in this study helped to lower the radiation dose to the ICP-OES operator and produced much less radioactive waste while at the same time improving LODs. Calibration curves were linear up to 4mg/L with a correlation coefficient of better than 0.999. Because of its robust and reproducible performance, no internal standard was required to correct for potential drifts in sensitivity of the ICP-OES. Three certified water reference materials (BCR®-617, LGC6177, SRM® 1643a) were employed to assess the accuracy of the developed ICP-OES procedure, the results of which underpin its reliability. In addition, Na and K concentrations calculated from emission intensities measured at two wavelengths confirmed that each dataset is consistent in itself. A comparison of sector field ICP-MS results of 23 samples originating from pyrochemical reprocessing with the data obtained applying the newly developed ICP-OES procedure revealed that the latter provides a reliable, straightforward and more robust analytical approach to determine Na, K, and Li in radioactive specimens.
ISSN:0026-265X
1095-9149
DOI:10.1016/j.microc.2012.03.018