An evaluation of three commercially prepared anion-exchange resin columns for separation of tetraiodothryonine in serum

Three different commercially prepared anion-exchange resin columns for thryoxine (T 4) separation from serum for the “T 4 by column” assay were evaluated using the protocol of the supplier. The claims by the suppliers for the distribution and recovery of T 4 in the first and second thyroxine-contain...

Full description

Saved in:
Bibliographic Details
Published inClinica chimica acta Vol. 69; no. 2; pp. 285 - 291
Main Authors Agnese, S.T., Shaw, W., Spierto, F.W.
Format Journal Article
LanguageEnglish
Published Netherlands Elsevier B.V 01.06.1976
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Three different commercially prepared anion-exchange resin columns for thryoxine (T 4) separation from serum for the “T 4 by column” assay were evaluated using the protocol of the supplier. The claims by the suppliers for the distribution and recovery of T 4 in the first and second thyroxine-containing eluates were experimentally evaluated by the addition of a tracer quantity of purified [ 125I]T 4. The average experimentally determined elution ratios of recoverable T 4 (sum of T 4 in first and second eluates) with these commercial columns were: Bio-Rad, 92.2 : 7.8 (90 : 10 claimed); Oxford, 87.8 : 12.2 (94 : 6 claimed); and Curtis Nuclear, 88.7 : 11.3 (80 : 20 claimed). The percent recoveries of T 4 in the first thyroxine-containing eluate were: Oxford, 82.3 + 9.70 ( x ± 1 S.D.); Bio-Rad, 91.75 ± 2.09; and Curtis Nuclear, 74.20 ± 6.14. Mean serum T 4 values obtained by the column method with all commercial columns tested were lower than competitive protein binding radioassay (CPBR) values if the former values were not corrected for recovery. When individual recovery correction factors were applied to column results, improved correlation and better correspondence of “T 4 by column” mean values with the CPBR values were noted. It is concluded that the largest part of the total variability of the “T 4 by column” assay is contributed by the Chromatographic step when the colorimetric step is performed with an automated technique.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0009-8981
1873-3492
DOI:10.1016/0009-8981(76)90506-4