Performance of a large Bragg-curve spectrometer

A large Bragg-curve spectrometer has been constructed and tested. The detector has a cylindrical geometry and operates with a homogeneous electric field. Energy resolutions of < 0.8% and Z resolutions of Z/ ΔZ = 80 have been achieved for elastically scattered 58Ni ions. These results demonstrate...

Full description

Saved in:
Bibliographic Details
Published inNuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Vol. 255; no. 3; pp. 507 - 511
Main Authors Vineyard, M.F., Wilkins, B.D., Henderson, D.J., Kovar, D.G., Beck, C., Davids, C.N., Kolata, J.J.
Format Journal Article
LanguageEnglish
Published Elsevier B.V 01.04.1987
Online AccessGet full text

Cover

Loading…
More Information
Summary:A large Bragg-curve spectrometer has been constructed and tested. The detector has a cylindrical geometry and operates with a homogeneous electric field. Energy resolutions of < 0.8% and Z resolutions of Z/ ΔZ = 80 have been achieved for elastically scattered 58Ni ions. These results demonstrate the suitability of this large solid-angle detector for use in a wide variety of heavy-ion scattering experiments.
ISSN:0168-9002
1872-9576
DOI:10.1016/0168-9002(87)91219-8