Design and Implementation of Cholecystolithiasis Automatically Perfusion-Dissolved Instrument

In view of the difficulty to operate, low efficiency, and easy to leak of the manual perfusion method for dissolving cholecystolithiasis, an automatically perfusion-dissolved instrument is designed. By means of collecting and analysing the pressure data being outputted from pressure sensor in real t...

Full description

Saved in:
Bibliographic Details
Published inApplied Mechanics and Materials Vol. 241-244; pp. 562 - 565
Main Authors Zhou, Tian Shui, Wang, Jun, Ling, Zhen Bao
Format Journal Article
LanguageEnglish
Published Zurich Trans Tech Publications Ltd 01.01.2013
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:In view of the difficulty to operate, low efficiency, and easy to leak of the manual perfusion method for dissolving cholecystolithiasis, an automatically perfusion-dissolved instrument is designed. By means of collecting and analysing the pressure data being outputted from pressure sensor in real time, and adjusting speed and direction of peristaltic pumps automatically to control quantity of dissolvent, so as to achieve the goal of litholysis under constant pressure in the gallbladder. The instrument has function of automatic heating and controlling temperature. Double controllers are used in the design, slave computer controls the running of each module of the instrument, and master computer is used to real-time display operation state and set parameters. Experimental results indicate that the system is stable and reliable, and with high efficiency for dissolving cholecystolithiasis, which provides a new and effective method for cholelithiasis medical treatment.
Bibliography:Selected, peer reviewed papers from the 2012 International Conference on Measurement, Instrumentation and Automation (ICMIA 2012), September 15-16, 2012, Guangzhou, China
ISBN:9783037855461
3037855460
ISSN:1660-9336
1662-7482
1662-7482