Fabrication of a metal-cored multi-barrelled microiontophoresis assembly
A method is described for fabrication of 7-barrelled microiontophoresis electrodes with a center barrel of platinum-coated tungsten. The electrodes require a minimum of expensive apparatus and can be fabricated in an hour or two. The electrodes have low recording impedance (typically 100 k omega and...
Saved in:
Published in | Journal of neuroscience methods Vol. 32; no. 1; p. 55 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
Netherlands
01.04.1990
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | A method is described for fabrication of 7-barrelled microiontophoresis electrodes with a center barrel of platinum-coated tungsten. The electrodes require a minimum of expensive apparatus and can be fabricated in an hour or two. The electrodes have low recording impedance (typically 100 k omega and low resistance iontophoresis barrels (typically 20-50 M omega). Compared to electrodes with a micropipette recording barrel, these electrodes are practically noise-free and can pass ionotophoretic currents of up to 200 nA without an appreciable increase in recording noise. |
---|---|
ISSN: | 0165-0270 |
DOI: | 10.1016/0165-0270(90)90071-M |