Magnetic resonance monitoring of stent deployment: in vitro evaluation of different stent designs and stent delivery systems

To evaluate MR imaging features of commercially available stents before, during, and after in vitro deployment as a step toward MR-guided stent deployment. Fourteen stents were deployed in a phantom under MR monitoring at 1.5 T by using a gradient-echo sequence. Device visibility was rated on a four...

Full description

Saved in:
Bibliographic Details
Published inInvestigative radiology Vol. 35; no. 6; p. 343
Main Authors Manke, C, Nitz, W R, Lenhart, M, Völk, M, Geissler, A, Feuerbach, S, Link, J
Format Journal Article
LanguageEnglish
Published United States 01.06.2000
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:To evaluate MR imaging features of commercially available stents before, during, and after in vitro deployment as a step toward MR-guided stent deployment. Fourteen stents were deployed in a phantom under MR monitoring at 1.5 T by using a gradient-echo sequence. Device visibility was rated on a four-point scale (excellent, fair, poor, not visible). The Memotherm stent and the rolling membrane (RM) Wallstent showed excellent stent visibility and at least fair scores for artifact-induced narrowing of the stent lumen. Three stents (Palmaz, AVE, Easy Wallstent) showed excellent visibility of the stent but no visible lumen. Five stents (Strecker, Accuflex, Hemobahn, Passager, Sinus) displayed fair visibility. The delivery catheters of four stent systems (Smart, Vascucoil, Symphony, ZA) displayed severe black hole artifacts. The imaging features of several stent systems might be suitable for MR-guided intervention. The Memotherm and the Wallstent RM combine good visibility of the stent and the lumen.
ISSN:0020-9996
DOI:10.1097/00004424-200006000-00002