Accuracy of a Real-Time Continuous Glucose Monitor in Pediatric Diabetic Ketoacidosis Admissions
Continuous glucose monitoring (CGM) devices are integral in the outpatient care of people with type 1 diabetes, although they lack inpatient labeling. Food and Drug Administration began allowing inpatient use during the coronavirus disease 2019 (COVID-19) pandemic, with some accuracy data now availa...
Saved in:
Published in | Diabetes technology & therapeutics Vol. 26; no. 9; p. 626 |
---|---|
Main Authors | , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
01.09.2024
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | Continuous glucose monitoring (CGM) devices are integral in the outpatient care of people with type 1 diabetes, although they lack inpatient labeling. Food and Drug Administration began allowing inpatient use during the coronavirus disease 2019 (COVID-19) pandemic, with some accuracy data now available, primarily from adult hospitals. Pediatric inpatient data remain limited, particularly during diabetic ketoacidosis (DKA) admissions and for patients receiving intravenous (IV) insulin.
This retrospective chart review compared point-of-care glucose values to personal Dexcom G6 sensor data during pediatric hospitalizations. Accuracy was assessed using mean absolute relative difference (MARD), Clarke Error Grids, and the percentage of values within 15/20/30% if glucose value >100 mg/dL and 15/20/30 mg/dL if glucose value ≤100 mg/dL.
Matched paired glucose values (
= 612) from 36 patients (median age 14 years, 58.3% non-Hispanic White, 47.2% male) and 42 inpatient encounters were included in this subanalysis of DKA admissions. The MARDs for DKA and non-DKA admissions (
= 503) were 11.8% and 11.7%, with 97.6% and 98.6% of pairs falling within A and B zones of the Clarke Error Grid, respectively. Severe DKA admissions (pH <7.15 and/or bicarbonate <5 mmol/L) had a MARD of 8.9% compared to 14.3% for nonsevere DKA admissions. The MARD during administration of IV insulin (
= 266) was 13.4%.
CGM accuracy is similar between DKA and non-DKA admissions and is maintained in severe DKA and during IV insulin administration, suggesting potential usability in pediatric hospitalizations. Further study on the feasibility of implementation of CGM in the hospital is needed. |
---|---|
ISSN: | 1557-8593 |
DOI: | 10.1089/dia.2023.0542 |