A low-cost long-range frequency domain low-coherence interferometry (LCI) detector is presented: time Fourier domain LCI (TFD-LCI). Combining ideas of time domain and frequency domain techniques, the TFD-LCI detects the analog Fourier transform of the optical interference signal with no limitation f...

Full description

Saved in:
Bibliographic Details
Published inJournal of the Optical Society of America. A, Optics, image science, and vision Vol. 40; no. 4; p. C16
Main Authors Cerrotta, Santiago, Torga, Jorge R, Morel, Eneas N
Format Journal Article
LanguageEnglish
Published United States 01.04.2023
Online AccessGet more information

Cover

Loading…
More Information
Summary:A low-cost long-range frequency domain low-coherence interferometry (LCI) detector is presented: time Fourier domain LCI (TFD-LCI). Combining ideas of time domain and frequency domain techniques, the TFD-LCI detects the analog Fourier transform of the optical interference signal with no limitation for the maximum optical path, measuring the thickness of several centimeters with micrometer resolution. A complete characterization of the technique is presented with a mathematical demonstration, simulations, and experimental results. An evaluation of repeatability and accuracy is also included. Measurements of small and large monolayer and multilayer thicknesses were done. Characterization of the internal and external thicknesses of industrial products such as transparent packages and glass windshield is presented, showing the potentiality of TFD-LCI for industrial applications.
ISSN:1520-8532
DOI:10.1364/JOSAA.478870