Optimal septum alignment of the Figulla Flex occluder to the atrial septum in patients with secundum atrial septal defects

To investigate the implantation safety, anatomic performance and septal alignment of the Occlutech Figulla Flex occluder (FFO) device, an atrial septal defect (ASD) closure device with specific left-sided deployment characteristics and right-sided septal alignment properties. Between January 2011 an...

Full description

Saved in:
Bibliographic Details
Published inEuroIntervention Vol. 11; no. 10; pp. 1153 - 1160
Main Authors Haas, Nikolaus A, Happel, Christoph M, Soetemann, Dagmar B, Hanslik, Andreas, Moysich, Axel, Kececioglu, Deniz, Laser, Kai Thorsten
Format Journal Article
LanguageEnglish
Published France 01.02.2016
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:To investigate the implantation safety, anatomic performance and septal alignment of the Occlutech Figulla Flex occluder (FFO) device, an atrial septal defect (ASD) closure device with specific left-sided deployment characteristics and right-sided septal alignment properties. Between January 2011 and December 2013 we prospectively collected the change of orientation of the device to the septum during the release process and the feasibility of implantation of the FFO in 122 patients. The mean age was 10.7 years (±10.2), weight 32.9 kg (±20.3), and height 129.4 cm (±30). Devices used were 9 (n=13), 10.5 (n=16), 12 (n=16), 15 (n=39), 18 (n=17), 21 (n=8), 24 (n=5), 27 (n=7) and 30 mm (n=3) in size. No additional implantation techniques were required. Before release, the mean angles of the left and right-sided discs were 29.2° (±9.9°) and 43.4° (±9.2°) to the body axis, and 18.7° (±8.7°) and 27.0° (±10°) immediately thereafter. Thus, there was only a slight change in orientation of the left-sided (10.6°±7.5°) and right-sided (16.3°±7.9°) discs. The design of this occluder system results in an ideal septum alignment which increases its feasibility as well as patient safety during implantation.
ISSN:1774-024X
1969-6213