Characteristic of an Organic Photodetector Fabricated With P3HT:ICBA Blending Materials for Indirect X-Ray Detection

In this study, an organic conjugated polymer-based photodetector was investigated as a candidate for the indirecttype radiation detector. In order to improve the photon to charge conversion efficiency, we selected the fullerene derivative indeneC60-bisadduct (ICBA) as an n-type semiconductor. The op...

Full description

Saved in:
Bibliographic Details
Published inIEEE transactions on nuclear science Vol. 64; no. 7; pp. 1739 - 1743
Main Authors Seon, Hyeji, Kim, Beomsu, Kang, Jungwon
Format Journal Article
LanguageEnglish
Published New York IEEE 01.07.2017
The Institute of Electrical and Electronics Engineers, Inc. (IEEE)
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:In this study, an organic conjugated polymer-based photodetector was investigated as a candidate for the indirecttype radiation detector. In order to improve the photon to charge conversion efficiency, we selected the fullerene derivative indeneC60-bisadduct (ICBA) as an n-type semiconductor. The optimal process condition was investigated while varying the P3HT:ICBA blending ratio and spin-rate. The detector at the condition of P3HT:ICBA = 3:2 blending ratio and 900 rpm spin-rate showed the highest sensitivity of 1.35 mC/Gy · cm 2 and the highest PCE of 2.93%. Compared to the common P3HT:PCBM detector, the best P3HT:ICBA detector showed 27.36% enhancement of sensitivity and 27.68% enhancement of the PCE.
ISSN:0018-9499
1558-1578