COMPARING TEE WITH ANGIOGRAPHIC SIZING OF WATCHMAN DEVICE FOR LEFT ATRIAL APPENDAGE CLOSURE

Overall agreement between TEE and angiographic measurement of LAA maximal width was 18% (kappa 0.13, p<0.01), moderate correlation (r=0.50; p<0.01); and agreement for LAA length was 7% (k= 0.03, p0.15) with modest correlation (r=0.33, p=0.01; figure) Agreement between hypothetical WATCHMAN siz...

Full description

Saved in:
Bibliographic Details
Published inJournal of the American College of Cardiology Vol. 69; no. 11; p. 1141
Main Authors Verma, Divya Ratan, Yassin, Haidar, Weiss, Jessica, Rabadi, Salem Al, Bhardwaj, Supriya, Morris, Michael, Gellert, George, Loli, Akil, Fang, H. Kenith, Byrne, Timothy, Pershad, Ashish
Format Journal Article
LanguageEnglish
Published New York Elsevier Inc 21.03.2017
Elsevier Limited
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Overall agreement between TEE and angiographic measurement of LAA maximal width was 18% (kappa 0.13, p<0.01), moderate correlation (r=0.50; p<0.01); and agreement for LAA length was 7% (k= 0.03, p0.15) with modest correlation (r=0.33, p=0.01; figure) Agreement between hypothetical WATCHMAN size vs actual WATCHMAN size implanted based on TEE was 19.6% (kappa 0.02, p=0.3); by angiogram was 17.7% (kappa 0.006, p=0.4), and integrated TEE + angiogram was 67.4% (kappa 0.6, p<0.001, Figure 1).
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
ISSN:0735-1097
1558-3597
DOI:10.1016/S0735-1097(17)34530-8