COMPARING TEE WITH ANGIOGRAPHIC SIZING OF WATCHMAN DEVICE FOR LEFT ATRIAL APPENDAGE CLOSURE
Overall agreement between TEE and angiographic measurement of LAA maximal width was 18% (kappa 0.13, p<0.01), moderate correlation (r=0.50; p<0.01); and agreement for LAA length was 7% (k= 0.03, p0.15) with modest correlation (r=0.33, p=0.01; figure) Agreement between hypothetical WATCHMAN siz...
Saved in:
Published in | Journal of the American College of Cardiology Vol. 69; no. 11; p. 1141 |
---|---|
Main Authors | , , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
New York
Elsevier Inc
21.03.2017
Elsevier Limited |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Overall agreement between TEE and angiographic measurement of LAA maximal width was 18% (kappa 0.13, p<0.01), moderate correlation (r=0.50; p<0.01); and agreement for LAA length was 7% (k= 0.03, p0.15) with modest correlation (r=0.33, p=0.01; figure) Agreement between hypothetical WATCHMAN size vs actual WATCHMAN size implanted based on TEE was 19.6% (kappa 0.02, p=0.3); by angiogram was 17.7% (kappa 0.006, p=0.4), and integrated TEE + angiogram was 67.4% (kappa 0.6, p<0.001, Figure 1). |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 |
ISSN: | 0735-1097 1558-3597 |