Polarization state generator: a polarimeter calibration standard
A polarization state generator (PSG) was built to generate light having a degree of linear polarization (DoLP) varying from 0.0005 to 0.4 with 0.0005 uncertainty. The PSG operates by tilting a plane parallel SF11 glass plate in a nearly unpolarized beam. The DoLP of collimated, unpolarized light pas...
Saved in:
Published in | Applied optics. Optical technology and biomedical optics Vol. 50; no. 12; p. 1726 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
United States
20.04.2011
|
Online Access | Get more information |
Cover
Loading…
Summary: | A polarization state generator (PSG) was built to generate light having a degree of linear polarization (DoLP) varying from 0.0005 to 0.4 with 0.0005 uncertainty. The PSG operates by tilting a plane parallel SF11 glass plate in a nearly unpolarized beam. The DoLP of collimated, unpolarized light passing through a plane parallel plate over a defined range of field angles can be calculated from theory, and the PSG was intended to act as a calibration standard based on this calculation. Several effects make the DoLP distribution as a function of field and tilt plate difficult to model to the desired 0.0005 uncertainty: source DoLP and intensity nonuniformity, lens surface diattenuation, and errors in optical alignment. Because of these effects, modeled DoLP was not used as a standard. Instead, DoLP was characterized with repeatability of 0.0005. |
---|---|
ISSN: | 2155-3165 |
DOI: | 10.1364/AO.50.001726 |