In vivo determination of nitrogen using Pu-Be sources
A mehtod of measuring body nitrogen by neutron capture gamma analysis is described. The neutron flux is provided by four collimated 5 Ci Pu-Be sources, placed to give a bilateral irradiation of the subject. The 10-83 MeV thermal neutron capture gamma rays from 14N are detected by two heavily shielde...
Saved in:
Published in | Physics in medicine & biology Vol. 22; no. 5; pp. 831 - 835 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
England
IOP Publishing
01.09.1977
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A mehtod of measuring body nitrogen by neutron capture gamma analysis is described. The neutron flux is provided by four collimated 5 Ci Pu-Be sources, placed to give a bilateral irradiation of the subject. The 10-83 MeV thermal neutron capture gamma rays from 14N are detected by two heavily shielded 5 in X 4 in NaI(Tl) detectors. Results indicate a reproducibility +/-3% (1 SD) for a 10 min irradiation of the chest area, while giving a radiation dose equivalent of 50 mREM. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0031-9155 1361-6560 |
DOI: | 10.1088/0031-9155/22/5/002 |