Measurements of gamma-ray background radiation in Pyhäsalmi mine

The results of measuring the gamma background radiation in Lab 5 of the underground facility in the Pyhäsalmi Mine at a depth of 1410 m are presented. The background integral count rate per kg of germanium of the HPGe-detector was 0.028 s-1 × kg-1. After purging the measuring chamber with nitrogen g...

Full description

Saved in:
Bibliographic Details
Published inApplied radiation and isotopes Vol. 161; p. 109166
Main Authors Pohuliai, S., Sokolov, A., Gostilo, V., Joutsenvaara, J., Puputti, J.
Format Journal Article
LanguageEnglish
Published England Elsevier Ltd 01.07.2020
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The results of measuring the gamma background radiation in Lab 5 of the underground facility in the Pyhäsalmi Mine at a depth of 1410 m are presented. The background integral count rate per kg of germanium of the HPGe-detector was 0.028 s-1 × kg-1. After purging the measuring chamber with nitrogen gas at a rate of 0.15 L/h the count rate was reduced to 0.021 s-1 × kg-1. •The gamma background radiation in Lab 5 in the Pyhäsalmi Mine are measured by Low background spectrometer.•Integral count rate, min. detectable activity were improved by purging measuring chamber of HPGe spectrometer.•The activities of natural radionuclides of the covering material in Lab 5 is significantly lower than that in Callio Lab.•The coating in the Callio Lab should be changed.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0969-8043
1872-9800
DOI:10.1016/j.apradiso.2020.109166