Deterioration of stannous ion in radiopharmaceutical kits during storage

The deterioration of stannous ion (Sn++) in inhouse-prepared and commercial radiopharmaceutical kits was studied. Sn++ content of three types of nonlyophilized, deoxygenated, aqueous inhouse-prepared kits [diethylenetriamine pentaacetic acid (DTPA), pyrophosphate and glucoheptonate] and of three com...

Full description

Saved in:
Bibliographic Details
Published inAmerican journal of hospital pharmacy Vol. 36; no. 10; p. 1370
Main Authors McBride, M H, Shaw, S M, Kessler, W V
Format Journal Article
LanguageEnglish
Published United States 01.10.1979
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:The deterioration of stannous ion (Sn++) in inhouse-prepared and commercial radiopharmaceutical kits was studied. Sn++ content of three types of nonlyophilized, deoxygenated, aqueous inhouse-prepared kits [diethylenetriamine pentaacetic acid (DTPA), pyrophosphate and glucoheptonate] and of three commercially prepared kits (two lyophilized pyrophosphate kits and one diphosphonate in sealed glass ampul kit) was measured by differential pulse polarography. Inhouse-prepared kits were assayed initially and after storage for 6, 12, 24 and 48 days at 24, 5 and -18 C. Commercial kits were assayed initially and after storage for 12, 24 and 48 days at 5 and 24 C. Of the inhouse-prepared kits, Sn++ stability when stored for 48 days at 5 and 24 C. Freezer storage should be used, when possible, to insure maximum stability of Sn++ in inhouse-prepared, nonlyophilized ratiopharmaceutical kits. The commercial procedures of lyophilization and of sealing the reagent in a sealed glass ampul prolong Sn++ stability.
ISSN:0002-9289
DOI:10.1093/ajhp/36.10.1370