A new titration technique is described in which the end-point is determined by measuring the relative reaction rate of the titration reaction. This technique is adequate for rather slow reactions where conventional direct titrations are not applicable. The titrations are done automatically under mic...

Full description

Saved in:
Bibliographic Details
Published inTalanta (Oxford) Vol. 30; no. 3; p. 145
Main Authors Efstathiou, C E, Hadjiioannou, T P
Format Journal Article
LanguageEnglish
Published Netherlands 01.03.1983
Online AccessGet more information

Cover

Loading…
More Information
Summary:A new titration technique is described in which the end-point is determined by measuring the relative reaction rate of the titration reaction. This technique is adequate for rather slow reactions where conventional direct titrations are not applicable. The titrations are done automatically under microcomputer control. The efficiency of this technique is demonstrated with direct titrations of certain polyhydroxy-compounds with standard periodate solution. Ethylene glycol and propylene glycol (0.05-0.3 mmole), glycerol (0.06-0.17 mmole) and mannitol (0.01-0.03 mmole) were determined with average relative errors of 0.1-0.3%.
ISSN:0039-9140
DOI:10.1016/0039-9140(83)80040-X