An Analytical Method to Calculate Phantom Scatter Factor for Photon Beam Accelerators
One of the important input factors in the commissioning of the radiotherapy treatment planning systems is the phantom scatter factor (Sp) which requires the same collimator opening for all radiation fields. In this study, we have proposed an analytical method to overcome this issue. The measurements...
Saved in:
Published in | Electronic physician Vol. 9; no. 1; pp. 3523 - 3528 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | English |
Published |
Iran
Electronic physician
01.01.2017
Electronic Physician |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | One of the important input factors in the commissioning of the radiotherapy treatment planning systems is the phantom scatter factor (Sp) which requires the same collimator opening for all radiation fields. In this study, we have proposed an analytical method to overcome this issue.
The measurements were performed using Siemens Primus Plus with photon energy 6 MV for field sizes from 5×5cm
to 40×40cm
. Phantom scatter factor was measured through the division of total scatter output factors (Scp), and collimator scatter factor (Sc).
The mean percent difference between the measured and calculated Sp was 1.00% and -3.11% for 5×5, 40×40 cm
field size respectively.
This method is applicable especially for small fields used in IMRT which, measuring collimator scatter factor is not reliable due to the lateral electron disequilibrium. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 2008-5842 2008-5842 |
DOI: | 10.19082/3523 |