A study on the characteristic of normoxic polymer gel dosimeter according to its composition

In this study, to find the optimal composition of the gel as therapeutic radiation, the amounts of methacrylic acid and gelatin were varied. The polymer gel with various compositions were evaluated for its sensitivity, reproducibility, and accuracy. As the concentration of the gelatine is high, the...

Full description

Saved in:
Bibliographic Details
Published inJournal of physics. Conference series Vol. 164; no. 1; p. 012018
Main Authors Cho, Sam Ju, Lim, Sangwook, Lee, Suk, Lee, Sang Hoon, Huh, Hyun Do, Cho, Kwang Hwan, Jang, Jisun, Choi, Jin Ho, Shin, Dong Oh, Kwon, Soo Il, Yun, Hyong Geun
Format Journal Article
LanguageEnglish
Published Bristol IOP Publishing 01.05.2009
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:In this study, to find the optimal composition of the gel as therapeutic radiation, the amounts of methacrylic acid and gelatin were varied. The polymer gel with various compositions were evaluated for its sensitivity, reproducibility, and accuracy. As the concentration of the gelatine is high, the threshold R2 value increases and the dose response was decreases. As the concentration of the methacrylic acid is high, both the threshold R2 value and the dose response were decrease. As both concentrations of the gelatine and the methacrylic acid is high, the sensitivity to the dose was increases within some range. It was found that the polymer gel composed in this study can be optimized for measuring the therapeutic radiation.