Impedance Spectroscopy Based on Microwave Probe Measurements
We have studied the influence of non-ideal standard loads of the probe station on the accuracy of microwave impedance spectrum measurements in the process of testing electronic devices and materials. Using the Cascade Microtech station as an example, it is shown that the calibration procedure based...
Saved in:
Published in | Radiophysics and quantum electronics Vol. 67; no. 7; pp. 546 - 557 |
---|---|
Main Author | |
Format | Journal Article |
Language | English |
Published |
New York
Springer US
01.12.2024
Springer Nature B.V |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | We have studied the influence of non-ideal standard loads of the probe station on the accuracy of microwave impedance spectrum measurements in the process of testing electronic devices and materials. Using the Cascade Microtech station as an example, it is shown that the calibration procedure based on standard loads leads to significant measurement errors. Prospects for improving load characteristics in order to eliminate errors are given. An original technique of impedance measurement, which is free of these problems, is considered. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 14 |
ISSN: | 0033-8443 1573-9120 |
DOI: | 10.1007/s11141-025-10396-1 |