Vertical root fracture diagnosis in teeth with metallic posts: Impact of metal artifact reduction and sharpening filters
This study examined the influence of a metal artifact reduction (MAR) tool, sharpening filters, and their combination on the diagnosis of vertical root fracture (VRF) in teeth with metallic posts using cone-beam computed tomography (CBCT). Twenty single-rooted human premolars - 9 with VRF and 11 wit...
Saved in:
Published in | Imaging science in dentistry Vol. 54; no. 2; pp. 139 - 145 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
Korea (South)
Korean Academy of Oral and Maxillofacial Radiology
01.06.2024
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | This study examined the influence of a metal artifact reduction (MAR) tool, sharpening filters, and their combination on the diagnosis of vertical root fracture (VRF) in teeth with metallic posts using cone-beam computed tomography (CBCT).
Twenty single-rooted human premolars - 9 with VRF and 11 without - were individually placed in a human mandible. A metallic post composed of a cobalt-chromium alloy was inserted into the root canal of each tooth. CBCT scans were then acquired under the following parameters: 8 mA, a 5×5 cm field of view, a voxel size of 0.085 mm, 90 kVp, and with MAR either enabled or disabled. Five oral and maxillofacial radiologists independently evaluated the CBCT exams under each MAR mode and across 3 sharpening filter conditions: no filter, Sharpen 1×, and Sharpen 2×. The diagnostic performance was quantified by the area under the receiver operating characteristic curve (AUC), sensitivity, and specificity. These metrics were compared using 2-way analysis of variance with a significance level of α=5%. Intra- and inter-examiner agreement were assessed using the weighted kappa test.
Neither MAR nor the application of sharpening filters significantly impacted AUC or specificity (
>0.05). However, sensitivity increased when MAR was combined with Sharpen 1× and Sharpen 2× (
=0.015). The intra-examiner agreement ranged from fair to substantial (0.34-0.66), while the inter-examiner agreement ranged from fair to moderate (0.27-0.41).
MAR in conjunction with sharpening filters improved VRF detection; therefore, their combined use is recommended in cases of suspected VRF. |
---|---|