A Picosecond Image-Converter Dissector with a Crossed Sweep

— A dual-slit dissector operating in the crossed-sweep mode with a picosecond time resolution has been created on the basis of the PIF-01 image converter tube. The dissector is applicable to synchrotron radiation sources and electron-positron colliders when it is necessary to simultaneously record l...

Full description

Saved in:
Bibliographic Details
Published inInstruments and experimental techniques (New York) Vol. 63; no. 1; pp. 87 - 91
Main Authors Andreev, S. V., Vorobiev, N. S., Dorokhov, V. L., Zarovsky, A. I., Komelkov, A. S., Kurkin, G. Ya, Meshkov, O. I., Smirnov, A. V.
Format Journal Article
LanguageEnglish
Published Moscow Pleiades Publishing 2020
Springer Nature B.V
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:— A dual-slit dissector operating in the crossed-sweep mode with a picosecond time resolution has been created on the basis of the PIF-01 image converter tube. The dissector is applicable to synchrotron radiation sources and electron-positron colliders when it is necessary to simultaneously record longitudinal profiles of multi-bunch beams that fill two to several hundred adjacent separatrices of an accelerator ring. Two trains of light pulses shifted in time with respect to each other have been recorded with a time resolution of 6.0 ± 0.5 ps during the dynamic tests of the developed dissector with a laser system.
ISSN:0020-4412
1608-3180
DOI:10.1134/S0020441220010017