Commissioning and Validation of a Dedicated Scanning Nozzle at Samsung Proton Therapy Center
In this study, we present the commissioning and validation results of a dedicated scanning nozzle. The dedicated scanning nozzle is installed in one of the two gantry treatment rooms at Samsung Proton Therapy Center. Following a successful completion of the acceptance test, the commissioning process...
Saved in:
Published in | Progress in Medical Physics Vol. 27; no. 4; pp. 267 - 271 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English Japanese |
Published |
Korean Society of Medical Physics
01.01.2016
한국의학물리학회 |
Subjects | |
Online Access | Get full text |
ISSN | 1226-5829 2508-4445 2288-9620 2508-4453 |
DOI | 10.14316/pmp.2016.27.4.267 |
Cover
Summary: | In this study, we present the commissioning and validation results of a dedicated scanning nozzle. The dedicated scanning nozzle is installed in one of the two gantry treatment rooms at Samsung Proton Therapy Center.
Following a successful completion of the acceptance test, the commissioning process including the beam data measurement for treatment planning system has been conducted. Extended measurements have been conducted as a validation of the clinical performance of the nozzle and various quality assurance protocols have been prepared. KCI Citation Count: 1 |
---|---|
Bibliography: | G704-001446.2016.27.4.008 |
ISSN: | 1226-5829 2508-4445 2288-9620 2508-4453 |
DOI: | 10.14316/pmp.2016.27.4.267 |