Evaluation of four techniques of warming intravenous fluids
This study was conducted to compare the fluid warming capabilities at different flow rates in four different warming systems. The intravenous (i.v.) fluid warmers used in this study were a water-bath warmer, a forced-air warmer, a dry-heat plate warmer, and an i.v. fluid tube warmer. Ringer's s...
Saved in:
Published in | Journal of anesthesia Vol. 16; no. 2; pp. 145 - 149 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
Japan
2002
|
Online Access | Get full text |
Cover
Loading…
Summary: | This study was conducted to compare the fluid warming capabilities at different flow rates in four different warming systems.
The intravenous (i.v.) fluid warmers used in this study were a water-bath warmer, a forced-air warmer, a dry-heat plate warmer, and an i.v. fluid tube warmer. Ringer's solution at room temperature or at ice-cold temperature (120-3000 ml.h(-1)) was used as an i.v. fluid, and the i.v. fluid temperatures were measured with thermocouples at the end of a 1-m tube connected to the warmer.
The temperature of the i.v. fluid delivered by the water bath and the forced-air warming system increased as the flow rate increased up to 1200 ml.h(-1) but decreased with further increase of the flow rate. The temperature of the i.v. fluid delivered by the dry-heat plate warmer significantly increased as the flow rate increased within the range studied. The delivered temperature did not depend on the flow rate within the range studied when the i.v. fluid tube warmer was used.
It is important to choose a warmer according to its characteristics as well as its performance. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0913-8668 1438-8359 |
DOI: | 10.1007/s005400200010 |