Permanent pacemakers: should straightened atrial leads be repositioned?

The aim of this study was to assess if atrial leads whose "J" configuration has straightened significantly on the postprocedural chest X ray should be repositioned. Between January 1996 and December 1997, 445 patients underwent dual chamber pacemaker implantation at the Papworth Hospital....

Full description

Saved in:
Bibliographic Details
Published inPacing and clinical electrophysiology Vol. 26; no. 11; pp. 2142 - 2145
Main Authors Hildick-Smith, David J R, Satchithananda, Duwarakan K, Newell, Stuart A, Grace, Andrew A, Murgatroyd, Francis D, Petch, Michael C
Format Journal Article
LanguageEnglish
Published United States 01.11.2003
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The aim of this study was to assess if atrial leads whose "J" configuration has straightened significantly on the postprocedural chest X ray should be repositioned. Between January 1996 and December 1997, 445 patients underwent dual chamber pacemaker implantation at the Papworth Hospital. Postprocedural chest X rays were available in 410 of these. The degree of straightening of the tip of the atrial lead was assessed from the lateral chest X ray and was graded as mild (-10 to +10 degrees from the horizontal), moderate (+10 to +30 degrees), or severe (> or = +30 degrees). Patients were followed with regard to atrial sensing and pacing characteristics, lead displacements, and lead revisions. Fifty-two (12%) patients had some degree of straightening (graded mild, moderate, severe) of the atrial lead on the postprocedure chest X ray (passive fixation in 48, active 4). Of these, 12 patients underwent next day lead repositioning, 5 of whom had abnormalities of pacing and/or sensing parameters. Seven patients therefore underwent repositioning of the atrial lead despite normal pacing parameters in view of lead straightening alone. Of the 12 patients who underwent repositioning, 3 still had lead straightening after the second procedure. The cohort for follow-up consisted of 43 patients (24 [56%] men, age 69 +/- 11 years at the time of implant) who were left with significant atrial lead straightening but adequate atrial parameters. Straightening was mild in 26 patients, moderate in 10, and severe in 7 patients. At implant the P wave amplitude was 4.8 +/- 2.4 mV. Follow-up was for 4.8 +/- 2.1 years, a total of 178 patient years. At final follow-up, the P wave amplitude was 2.7 +/- 1.3 (P < 0.05 vs implant). Censoring events occurred in 16 cases, comprising 11 deaths (none suspected to be pacemaker or lead related), 3 cases of persistent atrial fibrillation, 1 system extraction for infection, and 1 lead extraction for erosion. There were no cases of inadequate atrial lead sensing or pacing in the remaining patients. Irrespective of the degree of lead straightening on the postoperative lateral chest X ray, atrial leads should not be repositioned unless there are abnormalities of pacing or sensing parameters.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:0147-8389
1540-8159
DOI:10.1046/j.1460-9592.2003.00333.x