Speciation of Mn(II) and Mn(VII) by on-line spectrophotometric sequential injection analysis
An on-line sequential injection analysis system with spectrophotometric detection using 4-(2-pyridylazo) resorcinol as reagent is proposed for the speciation of Mn(II) and Mn(VII). The sampling frequency is 30 determinations per hour. The linear range for Mn(II) is 0.020–0.500 mg/l with a detection...
Saved in:
Published in | Analytica chimica acta Vol. 499; no. 1; pp. 129 - 137 |
---|---|
Main Authors | , , |
Format | Journal Article Conference Proceeding |
Language | English |
Published |
Amsterdam
Elsevier B.V
04.12.2003
Elsevier |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | An on-line sequential injection analysis system with spectrophotometric detection using 4-(2-pyridylazo) resorcinol as reagent is proposed for the speciation of Mn(II) and Mn(VII). The sampling frequency is 30 determinations per hour. The linear range for Mn(II) is 0.020–0.500
mg/l with a detection limit of 0.005
mg/l and for total Mn 0.025–0.550
mg/l with a detection limit of 0.008
mg/l. The % RSD is 0.27 and 0.34 for Mn(II) and Mn(VII), respectively (
n=10). Chemical masking of Fe and Al was achieved by on-line addition of a 0.1
mol/l NaF solution. The proposed system yielded results that compare very well with standard methods. |
---|---|
ISSN: | 0003-2670 1873-4324 |