What is the acceptable hemolysis index for the measurements of plasma potassium, LDH and AST?

Hemolysis is a cause of variability in test results for plasma potassium, LDH and AST and is a non-negligible part of measurement uncertainty. However, allowable levels of hemolysis provided by reagent suppliers take neither analytical variability (trueness and precision) nor the measurand into acco...

Full description

Saved in:
Bibliographic Details
Published inAnnales de biologie clinique (Paris) Vol. 74; no. 3; p. 323
Main Authors Rousseau, Nathalie, Pige, Raphaëlle, Cohen, Richard, Pecquet, Matthieu
Format Journal Article
LanguageEnglish
Published France 01.06.2016
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:Hemolysis is a cause of variability in test results for plasma potassium, LDH and AST and is a non-negligible part of measurement uncertainty. However, allowable levels of hemolysis provided by reagent suppliers take neither analytical variability (trueness and precision) nor the measurand into account. Using a calibration range of hemolysis, we measured the plasma concentrations of potassium, LDH and AST, and hemolysis indices with a Cobas C501 analyzer (Roche Diagnostics(®), Meylan, France). Based on the allowable total error (according to Ricós et al.) and the expanded measurement uncertainty equation we calculated the maximum allowable bias for two concentrations of each measurand. Finally, we determined the allowable hemolysis indices for all three measurands. We observed a linear relationship between the observed increases of concentration and hemolysis indices. The LDH measurement was the most sensitive to hemolysis, followed by AST and potassium measurements. The determination of the allowable hemolysis index depends on the targeted measurand, its concentration and the chosen level of requirement of allowable total error.
ISSN:1950-6112