Generation of reference electric and magnetic fields for calibration of power-frequency field meters

Improvements in the accuracy of power-frequency electric and magnetic field measurements require adequately calibrated instruments. The systems for the generation of reference electric and magnetic fields set up at the Istituto Elettrotecnico Nazionale 'Galileo Ferraris' (IEN) are describe...

Full description

Saved in:
Bibliographic Details
Published inIEEE transactions on instrumentation and measurement Vol. 42; no. 2; pp. 547 - 552
Main Authors Bottauscio, O., Crotti, G., D'Emilio, S., Farina, G., Mantini, A.
Format Journal Article Conference Proceeding
LanguageEnglish
Published New York, NY IEEE 01.04.1993
Institute of Electrical and Electronics Engineers
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Improvements in the accuracy of power-frequency electric and magnetic field measurements require adequately calibrated instruments. The systems for the generation of reference electric and magnetic fields set up at the Istituto Elettrotecnico Nazionale 'Galileo Ferraris' (IEN) are described. A description of the calibration procedures is given, together with a detailed uncertainty budget. The resulting uncertainty of the generated fields at the 1 sigma level is within 1.3*10/sup -3/ for the electric field up to 60 kV/m and within 6*10/sup -3/ for the magnetic field up to 10 mT.< >
Bibliography:ObjectType-Article-2
SourceType-Scholarly Journals-1
ObjectType-Feature-1
content type line 23
ISSN:0018-9456
1557-9662