In situ determination of 238Pu in the presence of uranium by triple quadrupole ICP-MS (ICP-QQQ-MS)
In this work, a direct determination of 238 Pu determination using reactive gases, CO 2 and H 2 in triple quadrupole ICP-MS without prior radiochemical separation resins has been established. Quantification of 0.25 pg/mL 238 Pu sample mixture (1.18 pg of 238 Pu) yielded satisfactory results (> 95...
Saved in:
Published in | Journal of radioanalytical and nuclear chemistry Vol. 322; no. 2; pp. 399 - 406 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | English |
Published |
Cham
Springer International Publishing
2019
Springer Nature B.V |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | In this work, a direct determination of
238
Pu determination using reactive gases, CO
2
and H
2
in triple quadrupole ICP-MS without prior radiochemical separation resins has been established. Quantification of 0.25 pg/mL
238
Pu sample mixture (1.18 pg of
238
Pu) yielded satisfactory results (> 95% accuracy), uncertainties of less than 15% and in the presence of spiked natural U (0.5 ng/mL). Additional verification of this method was performed using certified ratio materials (IRMM-086, Belgium) and experimental
238
Pu/
239
Pu and
240
Pu/
239
Pu values are analogous to the certified values provided. |
---|---|
ISSN: | 0236-5731 1588-2780 |
DOI: | 10.1007/s10967-019-06695-3 |