Automated indirect fluoride determination by ICP-OES using discontinuous-flow analysis
An automated method for the indirect determination of fluoride by inductively coupled plasma atomic emission spectrometry using discontinuous-flow analysis has been developed. A modified flow injection analysis system was used for sample preparation. The method is based on the determination of exces...
Saved in:
Published in | Toxicological and environmental chemistry Vol. 91; no. 7; pp. 1217 - 1227 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
Taylor & Francis Group
01.10.2009
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | An automated method for the indirect determination of fluoride by inductively coupled plasma atomic emission spectrometry using discontinuous-flow analysis has been developed. A modified flow injection analysis system was used for sample preparation. The method is based on the determination of excess cerium(III) ion after precipitation of CeF
3
. Separation of the precipitate is performed online on a syringe membrane filter that is cleaned each analysis cycle by 1 M nitric acid, ensuring a long life-time for the filter. As a result of the short reaction time, a detection limit of 9.5 mg L
−1
was obtained. The precision (RSD) of the proposed method is 1.02% at fluoride concentration of 28 mg L
−1
. The applicability of the technique for groundwater monitoring of fluoride-contaminated area is demonstrated. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0277-2248 1029-0486 |
DOI: | 10.1080/02772240802631907 |