Filters Effect on Image Quality of 177Lu-DOTATATE SPECT Images for Neuro-Endocrine Tumors Scanning

Purpose: The quality of the image in SPECT scans depends on the imaging parameters which are determined experimentally in the field of nuclear medicine. Designing a dedicated scanning parameter for 177Lu-DOTATATE SPECT images is required to optimize reconstruction. Therefore, this study aims to eval...

Full description

Saved in:
Bibliographic Details
Published inFrontiers in biomedical technologies Vol. 12; no. 1
Main Authors Dakhil, Hussein, Deevband, Mohammed Reza, Amoui, Mahasti, Mostaar, Ahmad, Ghorbani, Mahdi, Khosroabadi, Mohsen, Ghodsirad, Mohammad Ali
Format Journal Article
LanguageEnglish
Published Tehran University of Medical Sciences 01.12.2025
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Purpose: The quality of the image in SPECT scans depends on the imaging parameters which are determined experimentally in the field of nuclear medicine. Designing a dedicated scanning parameter for 177Lu-DOTATATE SPECT images is required to optimize reconstruction. Therefore, this study aims to evaluate the effect of different filters on image quality for bone SPECT scanning using 177Lu-DOTATATE. Materials and Methods: The filtered back-projection reconstruction method was used in neuro-endocrine tumor scanning using 177Lu-DOTATATE. Three hours after injection of 177Lu-DOTATATE, SPECT scans from 177Lu-DOTATATE for 30 patients were acquired using a dual-head EvoExel detector system. Seven parameters were considered, including the contrast/noise ratio, injection activity, uptake duration, acquisition time per injection, frame time, measuring time, and type of filters. Results: In all cases, the application of different filters increased Contrast to Noise Ratio (CNR) (9.1%, 1.8%, 10.9%, 61.8%, 23.6%, 29.1%, and 58.2% for Wiener, Butterworth, Parzen, Metz, Ramp, Shepp-Logan, and Hamming filters, respectively). The percentage of increase in Signal-to-Noise Ratio (SNR) was 3.3%, 1.7%, -24%, 21.7%, 9.8%, 11.9%, and 20.6%. Conclusion: Based on the quantitative analysis of the results, the application of the Metz filter (power 2) and Hamming filter (with 0.27, 0.47, and 0.67 cut-off frequencies) on SPECT scans of neuro-endocrine tumors is recommended because of their capacity to provide high-quality images.
ISSN:2345-5837
2345-5837
DOI:10.18502/fbt.v12i1.17727