Multichannel linear detector for x-ray spectroscopy
This paper considers a new linear multichannel x-ray detector designed on the basis of a BLPP-369M4 silicon photodiode array (2612 photodiodes, array pitch 12.5 µm, height 4 mm, and dynamic range 10 4 ). The structure and characteristics of the multichannel detector are given, along with the K α 1,2...
Saved in:
Published in | Optoelectronics, instrumentation, and data processing Vol. 45; no. 3; pp. 227 - 233 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
Heidelberg
Allerton Press, Inc
01.06.2009
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | This paper considers a new linear multichannel x-ray detector designed on the basis of a BLPP-369M4 silicon photodiode array (2612 photodiodes, array pitch 12.5 µm, height 4 mm, and dynamic range 10
4
). The structure and characteristics of the multichannel detector are given, along with the K
α
1,2, K
β
1
, and K
β
5
x-ray emission lines and a K-edge absorption spectrum of metallic copper recorded on a universal URS-2I spectrometer using this detector. The resolution and the signal/noise ratio of these spectra are superior to those of spectra recorded by an SRPP-21 gas ionization counter at the same recording time. The detector has a spatial resolution of 20 µm and an x-ray detection limit to equal 1 quantum at
λ
= 1.54 Å. It is suitable for studying the fine structure of absorption spectra at 1–10 Å. |
---|---|
ISSN: | 8756-6990 1934-7944 |
DOI: | 10.3103/S8756699009030029 |