Multichannel linear detector for x-ray spectroscopy

This paper considers a new linear multichannel x-ray detector designed on the basis of a BLPP-369M4 silicon photodiode array (2612 photodiodes, array pitch 12.5 µm, height 4 mm, and dynamic range 10 4 ). The structure and characteristics of the multichannel detector are given, along with the K α 1,2...

Full description

Saved in:
Bibliographic Details
Published inOptoelectronics, instrumentation, and data processing Vol. 45; no. 3; pp. 227 - 233
Main Authors Labusov, V. A., Mazalov, L. N., Fomenko, S. V., Selyunin, D. O., Bekhterev, A. V.
Format Journal Article
LanguageEnglish
Published Heidelberg Allerton Press, Inc 01.06.2009
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:This paper considers a new linear multichannel x-ray detector designed on the basis of a BLPP-369M4 silicon photodiode array (2612 photodiodes, array pitch 12.5 µm, height 4 mm, and dynamic range 10 4 ). The structure and characteristics of the multichannel detector are given, along with the K α 1,2, K β 1 , and K β 5 x-ray emission lines and a K-edge absorption spectrum of metallic copper recorded on a universal URS-2I spectrometer using this detector. The resolution and the signal/noise ratio of these spectra are superior to those of spectra recorded by an SRPP-21 gas ionization counter at the same recording time. The detector has a spatial resolution of 20 µm and an x-ray detection limit to equal 1 quantum at λ = 1.54 Å. It is suitable for studying the fine structure of absorption spectra at 1–10 Å.
ISSN:8756-6990
1934-7944
DOI:10.3103/S8756699009030029